(1888PressRelease)
March 12, 2007 - DESCRIPTION:
Fully automated random access continuous loading benchtop analyzer for Clinical Chemistry and Immunoturbidimetric assays.
ASSAY TIPE:
End Point, Kinetic, Differential, Bichromatic and Immunoturbidimetric assays; up to 100 different test, all user programmable.
TEST ENTRY MODE: selective - profiles- batch - STAT
ON LINE REAGENTS:
2 racks cooled by Peltier cells are provided to contain up to 20 reagents - 2 racks for 13 reagent positions at room temperature, 14 standards and controls - 2 more positions for diluents.
SAMPLES LOADING: 5 racks for continuous loading of samples split in:
64 positions for routine samples
14 positions for STAT samples
PRIMARY TUBES SIZE: max Æ 16 x 100
STAT SAMPLES PROCESSING: allowed at any time
THROUGHPUT: 200 test/hour without ISE
WORKING TEMPERATURE: 37°C
READING SYSTEM: direct photometry
MINIMUM REACTION VOLUME: 300 ml
OPTICS SYSTEM
PHOTOMETER: multi - wavelenght optics
WAVELENGHT: 8 narrow band interferential filters
LIGHT SOURCE: halogen lamp 6V/10W
OPTICAL PATH: 10 mm
LINEAR RANGE: 0.001 - 2.500 Abs
RESOLUTION: 0,001 Abs
SAMPLING ARM: one mechanic arm provides all sampling operation with: · capacitive liquid level sensing · reagent pre-warming at 37°C · automatic probe washing
DILUTER: dilutions are made without using traditional syringes with the following specifications:
sample volume: 20 µl - 50 µl (1 µl incr.)
reagent volume: 25 µl - 400 µl (1 µl incr.)
REACTION PLATE: circular 60 semi-disposable cuvettes tray with following carachteristics:
automatic cuvette washing
cuvettes Q.C. continuously computer controlled
incubation temperature: 37°C
COMPUTER & SOFTWARE-
COMPUTER TYPE: IBM compatible
MEMORY: 32 Mb
FD: 1.44 Mb · HD: 3.2 Gb
MONITOR: colour SVGA 15" low radiation
REPORTER PRINTER: 80 columns impact graphic
INTERFACE: bidirectional RS232C
SOFTWARE: multitasking WINDOWS 98
OPTIONS
ISE module
Positive barcode reader
Barcode printer
INSTALLATION REQUIREMENTS
POWER SUPPLY: 90 - 250 Vac; 47/63 Hz; 300W
WATER CONSUMPTION: approx 3 ml/test
DIMENSION, WEIGHT & ENVIRONMENT
ANALYTICAL MODULE: Height: 42 cm; Depht: 65 cm; Lenght: 100 cm; Weight 65 Kg AMBIENT TEMPERATURE: 18°C - 30°C
RELATIVE HUMIDITY: 20% - 85%